h2.Code 5273: «dmd blue fan — speed low» (Warning)
| Situation | Solution | |
| Wire unit of the fan disconnected. | Check if the wire unit of the fan is connected with the projector wiring. See reference 2 image. Check if the wiring of the fan is connected with the signal backplane. See reference 12 image. | |
| Blocked fan. | Unblock the fan. Ensure that the fan can turn freely. | |
| Damaged wire unit. | Check if the wire unit of the fan is not damaged. Repair if possible, otherwise replace with new one. | |
| Fan end of life. | Replace the fan. See «Replacement of the Formatter fan(s)». | |
| Malfunction Fan Control Board | Replace the Fan Control Board. See «Replacement of a Card Cage board». | |
| Other fans are also not spinning. | Replace the SMPS Board. See «Replacement of the SMPS board». | |
| Defect Signal Backplane | Replace the Signal Backplane. | |